14
If you will be using a Microtip
or Extender, remove the flat tip on the end of the horn,
then attach the Microtip
or Extender in its place.
Horns and probe tips wear after normal usage. Using a severely worn tip can damage
internal components. Replace tips when they become pitted. See the Maintenance
section for details.
For best results it is critical to use the appropriate size and type of horn. If you are not
sure that you have the proper horn, please call for assistance.
Low surface tension liquids (solvents) require a solid tip horn.
Set up regular preventive maintenance schedule. See the Maintenance section for
details.
Hearing protection is highly recommended. The Sound Enclosure (#432B) reduces
sonication noise to safe levels. If you decide not to use a Sound Enclosure, earplugs or
headphones should be worn.
If exposed to low temperatures the display may appear dim and difficult to read. Please
adjust the contrast to improve the view. Allowing the unit to warm to room temperature is
recommended.
In case of AC power loss, wait 3 minutes minimum before reapplying power.
Do not turn off AC mains power while running a horn. Stop sonication through the touch
screen prior to removing AC power.