Page 38 of 85 Operation
Programming an AC Hipot Test (continued)
END AC Hipot Test Programming. After selecting the FALL Time, one can either press [F1]
= NEW to change the step number in the highlighted box (1-10) and start programming STEP 3-
10, OR
one can press [F4] = EXIT to exit programming function and return to STAND BY
status.
For this example, press [F1] = NEW and proceed to program step 3 as a DC Hipot Test (¶ 2.6).
Set
Arc Limit
Set arc current limit:
1-20mA in 0.1mA increments
To move highlighted box to RAMP.
INC.
DEC.
EXIT
NEXT
RMT ERROFSTLOCK1 - 20mA 0 = OFF
F1
F4
F3
F2
FALL
RAMP
ARC
LOWACSTEP 2/2
VOLT
HIGH
TIME
INC.
DEC.
EXIT
NEXT
RMT ERROFSTLOCK0 - 999s 0 = OFF
F1
F4
F3
F2
FALL
RAMP
ARC
LOWACSTEP 2/2
VOLT
HIGH
TIME
INC.
DEC.
EXIT
NEXT
RMT ERROFSTLOCK0 - 999s 0 = OFF
F1
F4
F3
F2
FALL
RAMP
ARC
LOWACSTEP 2/2
VOLT
HIGH
TIME
To set Ramp Time:
0 - 999s in 0.1s increments
To move highlighted box to FALL.
Set
Ramp Time
To move highlighted box to STEP.
To set Fall Time:
0 - 999s in 0.1s increments
Set
Fall Time
NEW
MORE..
EXIT
NEXT
RMT ERROFSTLOCK1 - 10
F1
F4
F3
F2
FALL
RAMP
ARC
LOWACSTEP
VOLT
HIGH
TIME
2.750kV
To set next test step = 3
3.00mA
OFF
OFF
OFF
OFF
OFF
OFF
3.00mA
OFF
3.0s
OFF
3.00mA
3.0s
3.0s
OFF
3.00mA
To
exit
programming mode and return
to STAND BY status.
:
:
:
:
:
:
:
:
:
:
:
:
:
:
2.750kV
15.00mA
10.0s
2.750kV
15.00mA
10.0s
:
:
:
:
:
:
:
10.0s
2.750kV
15.00mA
:
:
:
:
:
:
:
10.0s
15.00mA
2/2
Set
Next
Test Step
[F1]
: Program Step 3
or
[F4] : Exit Program Mode