ONE O2 OXYGEN MONITOR ALARM UNITS Page 16 of 21.
6 APPENDIX A – TECHNICAL SPECIFICATION
MECHANICAL
(excl. cable) 82 (H) x 120 (W) x 56 (D) mm.
-5°C to +45°C (-20°C to +50°C at lower accuracy).
0 to 99% RH non-condensing
IP40 AS STANDARD
IP54 INSIDE W-PROOFBOX
Internal battery (part of ONEO2RS)
Standard Alarm Thresholds (Special thresholds available on request)
Low / Critically Low model
High / Critically High model
Response time (for correctly calibrated unit)
Typically less than 10 secs. for alarms to
activate when sensor subject to 18%
oxygen test gas
Status LED (yellow)
Low oxygen LED (red)
Critical oxygen LED (red)
CAL HIGH LED (green)
CAL LOW LED (green)
Battery LED (red)
Audible alarm (85dB)
Working Lifetime of the ONEO2 Sensor &
Battery Pack
The ONEO2 must be replaced by date
shown on label
RJ45 connection socket on top face for connection only to approved dedicated units.