Appendix D Calibration (Using Generator Service Tools)
HF Series X-ray Generators - Service Manual Revision W
Quantum Medical Imaging, LLC D-51
If the generator is equipped with "universal" type AEC (i.e., AEC
Board A11 is part number AY40-027S), and the ion chamber is
"level" type (i.e., Siemens), adjust potentiometer A11R82 to
increase or decrease the mAs value.
If the generator is equipped with "universal" type AEC (i.e., AEC
Board A11 is part number AY40-027S), and the ion chamber is
"slope" type (i.e., AID), adjust potentiometer A11R87 to increase
or decrease the mAs value.
10. Re-take exposure at the above settings. Repeat until the two mea-
sured values are within range.
11. Take an exposure at 110 kVp (100 kVp if calibrating a "level" type
(Siemens) ion chamber), 75S mA, 10 cm. Verify mAs value is within
±5% of previous measurements.
12. Take an exposure at 110 kVp (100 kVp if calibrating a "level" type
(Siemens) ion chamber), a mid-range mA value, 10 cm. Verify mAs
value is within ±5% of previous measurements.
CAUTION! If using a mAs meter, re-connect the
green/yellow wire marked "MAS METER" to the
ground stud on top of the H.V. Transformer (see
Figure D-21).