- 96 -
Command:
ARS_2
CR LF
set value release
parameter to
fast+reliable option
Response:
ARS_OK CR
LF
fast+reliable option
set
ARG Give current value release
Command overview:
On receipt of a command, the terminal sends information on currently value release.
Format:
ARG <CR><LF>
Response options:
ARG_x_OK<CR><LF>
- command carried out, returns set value release
ARG_I <CR><LF>
- command understood but not accessible at this moment
x - parameter, value release symbol
Example:
Command:
ARG<CR><LF>
- give current value release
Response:
ARG_2_OK<CR><LF>
- currently set value release parameter to fast+reliable
option
Set last digit
Format: LDS_n CR LF
Response options:
LDS_OK CR LF
command carried out
LDS_E CR LF
error in-course of command
execution, no parameter or incorrect
format
LDS_I CR LF
command understood but not
accessible at this moment
n parameter, decimal value determining last digit settings
n → 1 always
2 never
3 when stable
CAUTION!
The numbering is assigned to a particular last digit option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, last digit settings
are assigned to the working mode.
An example:
Command:
LDS_1 CR
LF
set last digit option
to value ‘always’
Response:
LDS_OK CR
LF
‘always’ value set
Cooperation with PUE 7.1, PUE 10 terminal
Format: NT CR LF
Response options:
ES CR LF
- command not understood (format
error)
MASS FRAME
- response: mass value in a basic
measuring unit
Response format:
1
2
3
4
5
6
7
8
9-18
19
20-22
23
24-32
33
34-36
37
38
39
40