- 95 -
FIS_OK CR LF
command carried out
FIS_E CR LF
error in-course of command
execution, no parameter or incorrect
format
FIS_I CR LF
command understood but not
accessible at this moment
n parameter, decimal value determining filter number
n → 1 very fast
2 fast
3 average
4 slow
5 very slow
CAUTION!
The numbering is assigned to a particular filter name and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, filter settings are
assigned to the working mode.
An example:
Command:
FIS_3 CR LF
set average filter
Response:
FIS_OK CR LF
average filter set
FIG Give current filter
Command overview:
On receipt of a command, the terminal sends information on currently set filter.
Format: FIG <CR><LF>
Response options:
FIG_x_OK<CR><LF>
- command carried out, returns set filter
FIG_I <CR><LF>
- command understood but not accessible at this moment
x
- parameter, filter symbols
Example:
Command:
FIG<CR><LF>
- give current filter
Response:
FIG_2_OK<CR><LF>
- currently set
average filter
Set Value release
Format: ARS_n CR LF
Response options:
ARS_OK CR LF
command carried out
ARS_E CR LF
error in-course of command
execution, no parameter or incorrect
format
ARS_I CR LF
command understood but not
accessible at this moment
n parameter, decimal value determining value release options
n → 1 fast
2 fast+reliable
3 reliable
CAUTION!
The numbering is assigned to a particular value release option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, value release
settings are assigned to the working mode
An example: