[SENSe:]ESPectrum<sb>:FILTer[:RRC]:ALPHa <Alpha>
This command defines the roll-off factor for the RRC filter.
The RRC filter is available if the power reference is the channel power.
Suffix: 
<sb>
.
1..n
Sub block in a Multi-SEM measurement
Parameters:
<Alpha> Range:  0  to  1
*RST:  0.22
Manual operation:  See " Alpha: " on page 230
[SENSe:]ESPectrum<sb>:FILTer[:RRC][:STATe] <State>
This command turns the RRC filter in the reference range on and off.
The RRC filter is available if the power reference is the channel power.
Suffix: 
<sb>
.
1..n
Sub block in a Multi-SEM measurement
Parameters:
<State> ON | OFF | 0 | 1
*RST:  1
Manual operation:  See " RRC Filter State " on page 230
[SENSe:]ESPectrum<sb>:RRANge?
This command queries the reference range.
Suffix: 
<sb>
.
Sub block in a Multi-SEM measurement
Return values: 
<RefRange> Number of the current reference range.
Range:  1  to  30
Usage:  Query only
[SENSe:]ESPectrum<sb>:RTYPe <Type>
This command defines the type of the power reference.
Suffix: 
<sb>
.
1..n
Sub block in a Multi-SEM measurement
Parameters:
<Type> PEAK | CPOWer