Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
927User Manual 1178.8520.02 ─ 01
If the probe is disconnected, the DM-mode offset of the probe is reset to 0.0 V.
Note that if the offset for DM-mode or CM-mode is changed, the offsets for the P-mode
and N-mode are adapted accordingly, and vice versa.
For details see the R&S FSV/A I/Q Analyzer and I/Q Input User Manual.
Suffix: 
<pb>
.
1..n
Selects the connector:
3 = RF
Parameters:
<DMOffset> Voltage offset between the positive and negative input terminal
Default unit: V
Manual operation:  See "Common Mode Offset / Diff. Mode Offset / P Offset / N Off-
set / " on page 315
[SENSe:]PROBe<pb>:SETup:MODE 
<Mode>
Suffix: 
<pb>
.
1..n
Selects the connector:
3 = RF
Parameters:
<Mode> RSINgle | NOACtion
RSINgle
Run single: starts one data acquisition.
NOACtion
Nothing is started on pressing the micro button.
Manual operation:  See " Microbutton Action " on page 316
[SENSe:]PROBe<pb>:SETup:NAME?
Queries the name of the probe.
Suffix: 
<pb>
.
1..n
Selects the connector:
3 = RF
Return values: 
<Name> String containing the name of the probe.
Example:  //Query name of the probe
PROB3:SET:NAME?
Usage:  Query only
Manual operation:  See "Name" on page 315
Setting Basic Measurement Parameters