Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1042User Manual 1178.8520.02 ─ 01
<dl> 1 to 4
frequency line
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example: 
CALC:FLIN2:STAT ON
Turns frequency line 2 on.
CALCulate<n>:TLINe<dl> 
<Time>
This command defines the position of a time line.
Suffix: 
<n>
.
Window
<dl> 1 to 4
time line
Parameters:
<Time> Note that you can not set a time line to a position that is higher
than the current sweep time.
Range:  0 s  to  1600 s
*RST:  (STATe to OFF)
Default unit: S
Example: 
CALC:TLIN 10ms
Sets the first time line to 10 ms.
Manual operation:  See "Vertical Line <x>" on page 469
CALCulate<n>:TLINe<dl>:STATe <State>
This command turns a time line on and off
Suffix: 
<n>
.
Window
<dl> 1 to 4
time line
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example: 
CALC:TLIN:STAT ON
Turns the first time line on.
Analyzing Measurements (Basics)