Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
984User Manual 1178.8520.02 ─ 01
<md> destination marker, see Marker
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example: 
CALC:DELT4:LINK:TO:MARK2 ON
Links the delta marker 4 to the marker 2.
Manual operation:  See " Linking to Another Marker " on page 431
CALCulate<n>:DELTamarker<m>:MODE <Mode>
This command defines whether the position of a delta marker is provided as an abso-
lute value or relative to a reference marker.
Note that when the position of a delta marker is queried, the result is always an abso-
lute value (see CALCulate<n>:DELTamarker<m>:X on page 985)!
Suffix: 
<n>
.
Window
<m> irrelevant
Parameters:
<Mode> ABSolute
Delta marker position in absolute terms.
RELative
Delta marker position in relation to a reference marker.
*RST:  RELative
Example: 
CALC:DELT:MODE ABS
Absolute delta marker position.
CALCulate<n>:DELTamarker<m>:MREFerence <Reference>
This command selects a reference marker for a delta marker other than marker 1.
Suffix: 
<n>
.
Window
<m> Marker
Parameters:
<Reference>
Example: 
CALC:DELT3:MREF 2
Specifies that the values of delta marker 3 are relative to marker
2.
Manual operation:  See " Reference Marker " on page 431
Analyzing Measurements (Basics)