FSV3000
982User Manual 1178.8520.02 ─ 01
CALC:SGR:FRAM:SEL -1
//Selects the frame that was captured 1 second after measurement start (Note the
//negative value!). This frame is displayed in the Spectrum window.
TRAC:DATA? SGR
//Retrieves the trace data for the spectrogram. For each frame, the power level
//and frequency at each sweep point are returned.
TRAC:DATA? TRACE1
//Retrieves the trace data for the selected frame only.
14.8.3 Working with Markers
The commands required to work with markers and marker functions in a remote envi-
ronment are described here. The tasks for manual operation are described in Chap-
ter 9.3, "Marker Usage", on page 426.
In the Spectrum application, markers are identical in all windows. Thus, the suffix <n>
for the window is generally irrelevant.
Setting Up Individual Markers............................................................................... 982
General Marker Settings....................................................................................... 988
Configuring and Performing a Marker Search.......................................................989
Positioning the Marker.......................................................................................... 993
Retrieving Marker Results.....................................................................................999
Marker Search (Spectrograms)...........................................................................1001
Fixed Reference Marker Settings........................................................................1010
Marker Peak Lists............................................................................................... 1012
Noise Measurement Marker................................................................................1015
Phase Noise Measurement Marker.....................................................................1017
Band Power Marker............................................................................................ 1019
n dB Down Marker.............................................................................................. 1023
Signal Count Marker........................................................................................... 1027
Marker Demodulation..........................................................................................1029
Programming Examples for Using Markers and Marker Functions.....................1031
14.8.3.1 Setting Up Individual Markers
The following commands define the position of markers in the diagram.
CALCulate<n>:DELTamarker<m>:AOFF...........................................................................983
CALCulate<n>:DELTamarker<m>:LINK............................................................................983
CALCulate<n>:DELTamarker<ms>:LINK:TO:MARKer<md>................................................983
CALCulate<n>:DELTamarker<m>:MODE......................................................................... 984
CALCulate<n>:DELTamarker<m>:MREFerence................................................................ 984
CALCulate<n>:DELTamarker<m>[:STATe]........................................................................ 985
CALCulate<n>:DELTamarker<m>:TRACe.........................................................................985
CALCulate<n>:DELTamarker<m>:X................................................................................. 985
CALCulate<n>:MARKer<m>:AOFF..................................................................................986
CALCulate<n>:MARKer<ms>:LINK:TO:MARKer<md>.......................................................986