Remote Control Commands
R&S
®
NRPxxS(N)
109User Manual 1177.5079.02 ─ 10
Manual operation: See "Frequency" on page 43
9.8.3 Setting Exclusion Time
In the burst average and the timeslot average modes, you can define a time period at
the beginning or at the end of an integration period, which is excluded from the mea-
surement result, as shown in Figure 9-6.
Figure 9-6: Effect of commands SENS:TIM:EXCL:STAR and SENS:TIM:EXCL:STOP
Remote commands:
[SENSe<Sensor>:]TIMing:EXCLude:STARt......................................................................109
[SENSe<Sensor>:]TIMing:EXCLude:STOP.......................................................................109
[SENSe<Sensor>:]TIMing:EXCLude:STARt <exclude_start>
Sets a time that is to be excluded at the beginning of the integration period, see Fig-
ure 9-6.
Parameters:
<exclude_start> Range: 0.0 to 1.0
*RST: 0.0
Default unit: Seconds
Manual operation: See "Start Exclude" on page 45
See "Start Exclude" on page 47
[SENSe<Sensor>:]TIMing:EXCLude:STOP <exclude_stop>
Sets a time that is to be excluded at the end of the integration period, see Figure 9-6.
Configuring Basic Measurement Parameters