3
- device RAT-2/RAT-Combo
- power transformer
- antenna
Before using the device, read this user manual.
Using this manual does not release you from the need to comply
with other safety regulations.
Any repairs or tampering with the device and its accessories may
only be performed by the manufacturer's authorized service
The device is covered by a 24 month warranty from the date of
purchase.
Due to the constant development of the device, the manufacturer
has the right to update this manual without having to inform the
users of the device.
The current version of the manual is available at www.ratmon.com
RAT-2 is the detector of leaks and faults in pre-insulated pipe network
Leak / Fault detection is done by measuring of 2 below parameters:
Rc - measurement of alarm loop continuity.
If Rc value exceeds the set limit of loop resistance it means loop’s break or misconnection of alarm
wires.
Riso - measurement of foam insulation resistance between metal pipe and alarm wire.
If the Riso value falls below the set insulation resistance limit it means
the appearance of moisture / leakage.
RAT-2 can support 2 or 4 alarm loops.
INTENDED USE OF THE DEVICE
CONTENT OF THE BOX