EasyManua.ls Logo

Raypak Temp-Tracker - Controller Output Testing

Raypak Temp-Tracker
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16
Connecting the Controller
Apply power to the Controller. The operation of the
Controller on power up is described in the Sequence
of Operation section, starting on page 4 of this manu-
al.
Testing the Controller’s
Outputs
The Controller has a built-in test routine, which is used
to test the main control functions. The test sequence is
enabled when the button is pressed and held for 3
seconds while in the View menu. The outputs are test-
ed in the following sequence:
1. After 1 second, the pump is turned on.
2. After 4 seconds, Stage 1 is turned on.
3. After 7 seconds, Stage 2 is turned on.
4. After 10 seconds, Stage 3 is turned on.
Fig. 22: Testing the Controllers Outputs
Sensor Resistance
Table D: Sensor Resistance
NOTE: Make sure ALL power to the devices and
wiring harness is off.
Fig. 21: Multi-meter
5
. After 13 seconds, Stage 4 is turned on.
6. After 16 seconds, the pump and stages 1 to 4 are
shut off. The alarm contacts are closed for 10 sec-
o
nds.
7
. The control exits the test sequence and resumes
normal operation.