42
Command is designed for rough calibration («zero installation») in the process of FLS
installation. If the length of probe is within (150…710) mm, «zero installation» is not required.
Command format:
Drift, byte
Value Description
0 1 31h Prefix
+1 1 00h…FFh
Network address of
acceptor
+2 1 45h Command code
+3 1 00h…3Fh
Value of parameter of
rough calibration
+4 1 00h…FFh CRC8
Answer format:
Drift, byte
Value Description
0 1 3Eh Prefix
+1 1 00h…FFh
Network address of
sender
+2 1 45h Command code
+3 1 00h or 01h
Return code: 00h -
command is complied
successfully, 01h -
Error
+4 1 00h…FFh CRC8
A2.3.2.2.2 Exact calibration (command 49h).
Command is designed for exact calibration («zero installation») in the process of FLS
installation. If the length of probe is within (150…710) mm, «zero installation» is not required
Command format:
Drift, byte
Value Description
0 1 31h Prefix
+1 1 00h…FFh
Network address of
acceptor
+2 1 49h Command code
+3 2 0000h…FFFFh
Value of parameter of
exact calibration
+5 1 00h…FFh CRC8
Answer format:
Drift, byte
Value Description
0 1 3Eh Prefix
+1 1 00h…FFh
Network address of
sender
+2 1 49h Command code
+3 1 00h or 01h
Return code: 00h -
command is complied
successfully, 01h -
Error