Instruction Manual DIGIMON-SE / DIGIMON4 English
19
9. Press the button to calibrate the high pressure side (HP) Repeat
from Point 6
10. Once both sensors are calibrated, press to exit the sensor calibration
mode.
Calibrating the temperature sensors / clamps (0°C / 32°F ice water)
1. Switch off the DIGIMON
2. Plug in the thermocouples (K-type).
The DIGIMON and the thermocouples must remain at a constant room
temperature for at least 20 minutes. The thermocouple cables must be straight
(unwound).
3. While holding down the button continuously, switch on the device by
pressing the button once. Continue holding down the Light button until
the display shows, "Password" 000
4. Press the / buttons to scroll to code number 020.
5. Press the button once.
Display shows “T-call0”
6. Press the button once and then press the button once to enter
temperature sensor/clamp calibration mode.
During calibration, ensure that both thermocouples are well connected to the
DIGIMON.
The display shows (low pressure side)
011xx
24.4°C
Temperature code (will change)
Your temperature reading (will not change)