EasyManua.ls Logo

ReliOn BP200W - Error Messages;Troubleshooting

ReliOn BP200W
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
If you do not intend to use the instrument for a week or more, always
remove the rechargeable batteries.
Recharge these batteries using an external charger and follow manu-
facturer’s instructions carefully.
7. Error messages/troubleshooting
If an error occurs during a measurement, the measurement is dis-
continued and a corresponding error code is displayed (Example:
Error No. 2).
Error No. Possible cause(s)
ERR 1 No pulse was detected.* Ensure the cuff is properly placed
on the inside of your wrist and repeat a measurement.
ERR 2 Unnatural pressure impulses influenced the measurement
result. Reason: The arm was moved during the measurement
(artefact). Repeat, keeping still and quiet.
ERR 3 Inflation of the cuff takes too long. The cuff is not correctly
seated. Re-position cuff and repeat the measurement.
ERR 5 The measured readings indicated an unacceptable difference
between systolic and diastolic pressures. Take another read-
ing following directions carefully. Contact your doctor if you
continue to get unusual readings.
HI The cuff pressure is too high. Relax for 5 minutes and repeat
the measurement.*
LO
The pulse is too low (less than 40). Repeat the measurement.*
* If this or any other problem occurs repeatedly, please consult
your doctor.
15

Related product manuals