Use of test fixture is mandatory during DSA6000 final HI testing. The HI fixture/jig is to be placed in the black test frame. Place the HI in the fixture
according to pic and place it all in the test chamber and select the correct test tpix and make sure to follow instructions coming up on the screen.
During the test programme execution the IAD calibration will be performed at the appropriate time. Another benefit from the fixture is that the HI
is in a well defined position so the telecoil test can be performed without any interaction from the operator
HI Testing - LO85 Models
The test tubing is normal 20 mm eective
lenght as for all BTE testing
20 mm
2cc coupler
p/n 30-4838800
2cc coupler insert BTE
p/n 40B4609601
Coupl Mic,TM-12,measured
p/n 18220700