1
Amara View minimal contact full-face mask
Instructions for Use
Intended Use
This mask is intended to provide an interface for application of CPAP or bi-level therapy
to patients. The mask is for single patient use in the home or multi-patient use in the
hospital/institutional environment. The mask is to be used on patients (>66lbs/30kg) for
whom CPAP and bi-level therapy has been prescribed.
Caution: U.S. Federal law restricts this device to sale by or on the order of a
physician.
Note:
An exhalation port is built into this mask so a separate exhalation port is not required.
This mask is not made with natural rubber latex or DEHP.
Symbols
Warning or Caution
Note
Tip
Consult Instructions for
Use
Not Made with Natural
Rubber Latex
X1 System One Resistance Control
Warnings
This mask is not suitable for providing life support ventilation.
This mask is designed for use with CPAP or bi-level systems recommended by your
health care professional or respiratory therapist. Do not wear this mask unless the CPAP
or bi-level system is turned on and operating properly. Do not block or try to seal the
exhalation port. Explanation of the Warning: CPAP systems are intended to be used
with special masks with connectors which have vent holes to allow continuous flow
of air out of the mask. When the CPAP machine is turned on and functioning properly,
new air from the CPAP machine flushes the exhaled air out through the attached
mask exhalation port. However, when the CPAP machine is not operating, enough
fresh air will not be provided through the mask, and exhaled air may be rebreathed.
Rebreathing of exhaled air for longer than several minutes can in some circumstances
lead to suffocation. This warning applies to most models of CPAP systems.
Question and Answer IconNeed help?

Do you have a question about the Respironics Amara View and is the answer not in the manual?

Respironics Amara View Specifications

General IconGeneral
BrandRespironics
ModelAmara View
CategoryRespiratory Product
LanguageEnglish

Summary

Intended Use

Symbols

Warnings

Contraindications

Before Use Read and Understand the Instructions Completely.

Cleaning Instructions

Institutional Disinfection

Achieving the Right Fit

Sizing the Mask

Guidance on correctly sizing the mask to ensure comfort and proper seal against the face.

Before putting on the mask:

Steps to prepare the mask and headgear before applying it to the face.

Adjusting the mask:

Detailed steps for fitting and adjusting the mask for optimal comfort and seal.

Using the mask:

Comfort Tips

Removing the Mask

Disassembly

Assembly

Philips Respironics System One Resistance Control

Specifications

Pressure Flow Curve

Graphical representation of the mask's flow rate versus pressure, indicating performance characteristics.

Resistance with Anti-Asphyxia Valve Closed to Atmosphere

Specifies pressure drop values for different flow rates when the anti-asphyxia valve is closed.

Closed to Atmosphere

Requirement for maintaining a minimum pressure to ensure proper closure of the anti-asphyxia valve.

Open to Atmosphere

Describes the behavior of the anti-asphyxia valve when positive airway pressure is absent.

Inspiratory and Expiratory Resistance with Anti-Asphyxia Valve Open to Atmosphere at 50 SLPM

Details inspiratory and expiratory resistance values at a specific flow rate with the valve open.

Deadspace

Sound Levels

Disposal

Storage Conditions

LIMITED WARRANTY