What to do if my Respironics EverGo Oxygen Equipment won't trigger a pulse of oxygen?
- RRachel SilvaAug 1, 2025
If your Respironics Oxygen Equipment isn't delivering a pulse of oxygen, first ensure the device is turned on by touching the blank screen and then the On/Off symbol. Check that the cannula tubing is properly connected to the oxygen outlet port and free of kinks or obstructions. If this doesn't resolve the issue, allow the device to run for 2 minutes without breathing through the cannula before trying again. If the issue persists, it may be a device malfunction.