25
sense of EN60601-1, which would present an unacceptable risk to patients, operators or
third parties should the power supply fail or malfunction.
Warning:
Portable RF communications equipment (radios) including accessories, such as antenna
cables and external antennas, should not be used in closer proximity than 30 cm (12 in-
ches) to parts and cables of the ri-magic®HPLED specified by the manufacturer. Failure to
comply may result in reduced device performance.
Guidelines and manufacturer‘s declaration – electromagnetic emissions
The ri-magic HPLED examination lamp is intended for use in an electromagnetic en-
vironment as specified below. The customer or user of the headlight should ensure that
it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Groupe 1
The ri-magic® HPLED examination lamp
only uses RF energy for its internal functi-
oning. Therefore, its HF emissions are very
low and are unlikely to cause interference
in nearby electronic devices.
RF emissions
CISPR 11
Class A
The ri-magic® HPLED examination lamp
is suitable for use in all establishments,
including domestic establishments and
those directly connected to the public
low-voltage network that supplies buildings
for domestic purposes.
Harmonic emis-
sions
IEC 61000-3-2
Class A
Voltage fluctu-
ations/flicker
emissions
IEC 61000-3-3
Compliant