Perform the following maintenance tasks at regular intervals to maintain the
performance
of
SmartLab SE.
I Optics and detector maintenance
Multilayer mirror
Re-adjustment
is required
after performing
maintenance
of
the X
-ra
y generator.
Attenuator
Re-adjustment
is
required after changing the
correction
wavelength
or
performing maintenance
of
the
X-
ray
generator.
Detector Re-adjustment is required after changing the
wavelength to be measured.
There is a mathematical possibility that the characteristics
of
the detector wi
ll
change over time. However, you can minimize performance degradation by
making the appropriate adjustments. We recommend performing detector
adjustments at periodic intervals.
D/t
eX
Adjustment
HyPi
xA
djustment
HyPix
Ca
librati
on
To make the
se
adjustments automatically,
se
lect Package Part panel and
Activities panel, then
dra
g and drop the appropriate package activities
on
the
Sequence.
For
details about the
se
parts activities, refer to the Help topic
of
eac
h
Parts Activities.
SmartLab SE: Automated Multipurpose X-ray Diffractomet
er
105