What to do if the X-ray generator stops during measurement in Rigaku Laboratory Equipment?
M
Marcus GonzalezAug 7, 2025
If the X-ray generator stops during measurement or other operations, go to the Utility tab in SmartLab Studio II and click on System Status. The part name where the alarm has been generated will be displayed in red. Check the cause to resolve the issue.
P
Patrick Shelton Jr.Aug 13, 2025
What to do if the goniometer stops abruptly during measurement with Rigaku Laboratory Equipment?
T
Todd CampbellAug 13, 2025
If the goniometer stops abruptly during measurement or other operation, it may be due to excessive force applied to the theta_ s or theta_ d arm. Turn the power key on the front panel to OFF to stop SmartLab SE. Check the arms to see if any item obstructs their operation, then restart SmartLab SE.
T
Terri LucasAug 19, 2025
How to reset alarm status on Rigaku Laboratory Equipment?
J
James ArcherAug 19, 2025
To reset the alarm status, click 'Alarm' in the Utility tab, and then click 'Reset'.
L
Laura MoralesAug 26, 2025
What to do if an error occurs during SmartLab SE startup of Rigaku Laboratory Equipment?
F
fperezAug 26, 2025
If an error occurs during SmartLab SE startup, it could indicate a connection problem between SmartLab SE and the controller. First, check that the Power-On indicator is lit and the Alarm indicator is off. Also, confirm that the connectors are properly connected. After verifying these items, shut down SmartLab Studio II, and then restart it. When restarting SmartLab SE, make sure to restart SmartLab as well.
N
Nicole HallAug 31, 2025
How to open the door of SmartLab SE Rigaku Laboratory Equipment?
D
Darren FryeAug 31, 2025
To open the door of SmartLab SE, first confirm that the Door-Lock indicator is lighting up in green. Then, click the door lock button of the Smart Lab Studio II to unlock the door.