Rigaku Smartlab SE - Alarm Indicator and Alarm Sound Operations

Default Icon
117 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table Causes
of
alarm
Cause Description
XG error
Attachment connecting
cover
is
open
Communication error
Rigaku
An error has occurred
in
the X-ray generator (XG).
See#
4
in
Troubleshooting.
The cover
of
the attachment connection cable
is
open
in
the radiation enclosure.
Close the cover.
A communication error has occurred
in
the control board for the optical devices
in the incident and receiving optics. The cables for the incident and receiving
optics may be disconnected
or
broken.
SmartLab SE: Automated Multipurpose X-ray Diffractometer