RIGOL  Chapter 2 Command System 
2-154  DG800 Programming Guide 
:SYSTem:PRESet:SAVe 
 
Syntax 
:SYSTem:PRESet:SAVe {USER1|USER2|USER3|USER4|USER5|USER6|USER7|USER8|USER9|USER10} 
 
Description 
Saves the current system status to the specified user storage location (USER1 to USER10) in the internal 
memory of the instrument. 
 
Parameter 
{USER1|USER2|USER3|USER4|
USER5|USER6|USER7|USER8|
Discrete 
USER1|USER2|USER3|USER4|
USER5|USER6|USER7|USER8|
None 
 
Example 
:SYSTem:PRESet:SAVe USER1 /*Saves the system status to the user storage location 1 in the internal 
memory of the instrument.*/ 
 
 
:SYSTem:PRESet[:STATe]? 
 
Syntax 
:SYSTem:PRESet[:STATe]? 
 
Description 
Queries user-defined 10 system status names. 
 
Return Format 
The query returns 10 system status names. If User n saves the current system status, then the query 
returns the current system status name. For example, "Default","","User2","","","","","","","","". 
 
 
:SYSTem:ROSCillator:SOURce 
 
Syntax 
:SYSTem:ROSCillator:SOURce {INTernal|EXTernal} 
:SYSTem:ROSCillator:SOURce? 
 
Description 
Sets the system clock source to internal source (INTernal) or external source (EXTernal). 
Queries the system clock source type. 
 
Parameter 
 
Remarks 
  DG800 provides an internal 10 MHz clock source and receives the external clock source from the 
rear-panel [10MHz In/Out] connector. It can also output clock source from the [10MHz In/Out] 
connector for other devices. 
  If "External" is selected, the instrument will detect whether a valid external clock signal is input from 
the rear-panel [10MHz In/Out] connector. If no valid clock source is detected, a prompt message is 
displayed, and the clock source will be switched to Internal. 
  You can synchronize two or multiple instruments by setting the clock source. 
  Synchronization between two instruments: