RIGOL Chapter 2 Command System
2-54 DG800 Programming Guide
Remarks
When [:SOURce[<n>]] or [<n>] is omitted, the system sets the related parameters of CH1 by default.
The range of <amp> is limited by the "Impedance" (
:OUTPut[<n>]:IMPedance
or
:OUTPut[<n>]:LOAD) setting. The range of <offset> is limited by the "Impedance" and
"Amplitude/High Level" settings.
Example
:SOUR1:APPL:NOIS 1,2 /*Sets the waveform of CH1 to Noise with 1 Vpp amplitude and 2 V
dc
offset.*/
[:SOURce[<n>]]:APPLy:PRBS
Syntax
[:SOURce[<n>]]:APPLy:PRBS
[<freq>|DEFault|MINimum|MAXimum[,<amp>|DEFault|MINimum|MAXimum[,<offset>|DEFault|MINimu
m|MAXimum]]]
Description
Sets the waveform of the specified channel to PRBS waveform with the specified frequency, amplitude, and
offset.
Parameter
Name
Type
Range
Default
[<n>]
Discrete
1|2
1
<freq>
Real
2 kbps to 30 Mbps
2 kbps
<amp>
Real
Refer to "Remarks".
5 Vpp
<offset>
Real
Refer to "Remarks".
0 V
dc
Remarks
When [:SOURce[<n>]] or [<n>] is omitted, the system sets the related parameters of CH1 by default.
The range of <amp> is limited by the "Impedance" (
:OUTPut[<n>]:IMPedance
or
:OUTPut[<n>]:LOAD) and "Frequency/Period" settings. The range of <offset> is limited by the
"Impedance" and "Amplitude/High Level" settings.
Example
:SOUR1:APPL:PRBS 10000,1,2 /*Sets the waveform of CH1 to PRBS with 10 kHz frequency, 1 Vpp
amplitude, and 2 V
dc
offset.*/
[:SOURce[<n>]]:APPLy:PULSe
Syntax
[:SOURce[<n>]]:APPLy:PULSe
[<freq>|DEFault|MINimum|MAXimum[,<amp>|DEFault|MINimum|MAXimum[,<offset>|DEFault|MINimu
m|MAXimum[,<phase>|DEFault|MINimum|MAXimum]]]]
Description
Sets the waveform of the specified channel to Pulse with the specified frequency, amplitude, offset, and
phase.
Parameter
Name
Type
Range
Default
[<n>]
Discrete
1|2
1
<freq>
Real
1 μHz to 10 MHz
1 kHz
<amp>
Real
Refer to "Remarks".
5 Vpp
<offset>
Real
Refer to "Remarks".
0 V
dc
<phase>
Real
0° to 360°

Other manuals for Rigol DG822