EasyManua.ls Logo

Rigol DM3061

Rigol DM3061
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL
Service Guide for DM3000 Series
3-4
1. Zero Offset Test
When testing Zero Offset, firstly apply a 4-wire short across the Input HI-Lo and
Sense HI-Lo terminals (front terminals), then leave the current terminal open as
shown in figure 3-1:
Figure 3-1
The Input HI-LO and Sense/Ref HI-LO Terminals in Short Circuit
Next, perform the test items in table 3-2 or table 3-3 one by one, then compare
the test results with the error range in the tables.