EasyManua.ls Logo

Rigol DM3061

Rigol DM3061
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL
Service Guide for DM3000 Series
3-14
Optional AC Voltage Test
In optional AC voltage test, input the standard signal the test equipment outputs into
the input terminal of tested equipment; and compare the test results with the error
shown in the table 3-12 and 3-13.
Table 3-12 Accessional AC Voltage Test Error (DM306x)
Test Signal
Input Frequency
Range
Error (1 year)
2V
10Hz
2.000000V
± 103.4mV
2V
1kHz
2.000000V
± 3.6mV
2V
50kHz
2.000000V
± 9.4mV
2V
100kHz
2.000000V
±14mV
2V
300kHz
2.000000V
±96mV
20V
1kHz
20.00000V
± 36mV
200V
1kHz
200.0000V
± 360mV
200mV
1kHz
200.0000mV
± 440μV
Notes:
[1] Set the AC filter as slow ( Meas Filter Slow );
[2] Set the reading resolution as .
Table 3-13 Accessional AC Voltage Test ErrorDM305X
Test Signal
Input Frequency
Range
Error (1 year)
2V
10Hz
2.000000V
± 22mV
2V
1kHz
2.000000V
± 6mV
2V
50kHz
2.000000V
± 22mV
2V
100kHz
2.000000V
± 44mV
20V
1kHz
20.00000V
± 60mV
200V
1kHz
200.0000V
± 600mV
200mV
1kHz
200.0000mV
± 600μV
Notes:
[1] Set the AC filter as slow ( Meas Filter Slow );
[2] Set the reading resolution as .

Other manuals for Rigol DM3061

Related product manuals