EasyManua.ls Logo

Rigol DM3061 - Page 48

Rigol DM3061
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL
Service Guide for DM3000 Series
3-6
Table 3-3 Zero Offset of the Standard Equipment (DM305X)
Input
Function
Quick Test
Point
Range
Error (1 year)
Open
DC Current
2.000000mA
± 1.4μA
Q
20.00000mA
± 1.6μA
200.0000 mA
± 18μA
1.000000A
± 700μA
10.00000A
[1]
± 700μA
Short
DC Voltage
Q
400.0000mV
± 32μV
4.000000V
± 240μV
40.00000V
± 2.4mV
400.0000V
± 24mV
1000.000V
[2]
± 50mV
Short
2-Wire
Ohms/
4-Wire Ohms
[3]
400.0000Ω
± 40mΩ
4.000000kΩ
± 240mΩ
40.00000 kΩ
± 2.4Ω
Q
400.0000 kΩ
± 28Ω
4.000000MΩ
± 400Ω
100.00000MΩ
± 5kΩ
Notes:
[1] Continuous current more than DC 7A or AC RMS 7A should be 30 seconds Off after 30 seconds
On;
[2] Each additional volt over ± 500 VDC adds 0.02 mV of error;
[3] Errors in above table are corresponding to 2-wire resistance function in math Null. Without
Math Null, the error is ± 0.2 Ω additional errors in 2–wire resistance function;
[4] The reading resolution of multimeter is 5½ ;
[5] Q: Optional quick test point.

Other manuals for Rigol DM3061

Related product manuals