Chapter 2 Front Panel Operation RIGOL
User’s Guide for DSA800/E Series
2-29
Auto SWT
Set the auto setting method of the sweep time to Normalor Accy. The sweep
will be faster if “Normal is selected; while in Accytype, higher measurement
precision can be obtained.
Mode
Set the sweep mode to Single or Cont. The default is Cont. The corresponding
icon of the sweep mode will be displayed in the status bar at the left side of the
screen.
Single Continuous
1. Single
Set the sweep mode to Single. The number 10 on the parameter icon
denotes the current sweep number.
2. Cont
Set the sweep mode to Cont. The character Cont on the parameter icon
denotes the analyzer is sweeping continuously.
Key Points:
If the instrument is in single sweep mode and no measurement function is
enabled, press this key and the system will enter continuous sweep mode and
sweep continuously if the trigger conditions are satisfied.
If the instrument is in single sweep mode and a measurement function is on,
press this key and the system will enter continuous sweep mode and measure
continuously if the trigger conditions are satisfied.
In continuous sweep mode, the system will send a trigger initialization signal
automatically and enter the trigger condition judgment directly after each
sweep.

Table of Contents

Other manuals for Rigol DSA832

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DSA832 and is the answer not in the manual?

Rigol DSA832 Specifications

General IconGeneral
BrandRigol
ModelDSA832
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals