RIKEN PORTABLE HC/O2 DETECTOR
MODEL RX−415
OPERATION MANUAL
(PT0-028)
Precautioninoperation

After reading this manual well, start to operate this instrument.
Keep this manual where to carry out easily
When take out this manual by use of construction work, be sure to put it
where it was.
Do not use this instrument for no other than given purpose.
When operate this instrument without using this manual or repair it by use of
no other parts than the genuine one, the safety and quality of products could
not be guaranteed. When any accident should take place by those irregular
method, we cannot assume the responsibility for it.
2-7-6 Azusawa, Itabashi-ku, Tokyo, 174-8744, Japan
Phone : +81-3-3966-1113
Fax : +81-3-3558-9110
E-mail : intdept@rikenkeiki.co.jp
Web site : http://www.rikenkeiki.co.jp/english/

Questions and Answers

  • N
    Nicholas NguyenAug 2, 2025
    What to do if zero gas display is flickering on Riken Keiki RX-415 Measuring Instruments?
    • A
      Amber HornAug 3, 2025
      If the “0.0”%LEL gas display is flickering on your Riken Keiki Measuring Instruments, the zero point may be deviated. Ensure the surrounding area is free of gas and then perform a zero adjustment.
  • M
    michaelbrockAug 8, 2025
    Why is “100.0vol%” flickering on my Riken Keiki RX-415 Measuring Instruments?
    • G
      Gregory AndersonAug 8, 2025
      If you observe a flickering “100.0vol%” on your Riken Keiki Measuring Instruments, it indicates that the zero point or gas sensitivity is deviated. First, perform a zero adjustment. If the reading value is still not adjusted, proceed with gas calibration.
  • C
    Christopher CarrollAug 15, 2025
    What to do if there is no pump suction in Riken Keiki RX-415 Measuring Instruments?
    • R
      robinsonmarkAug 15, 2025
      If your Riken Keiki Measuring Instruments has no pump suction, the pipe of IN, OUT of filter or tube may be removed or clogged. Check for any clogs or twists in the filter pipe and correct them.
  • K
    Katelyn WilsonAug 20, 2025
    Why is zero and span adjustment for O2 not available on Riken Keiki Measuring Instruments?
    • C
      Cassandra CamachoAug 20, 2025
      If zero and span adjustment for O2 is not available on your Riken Keiki Measuring Instruments, it may be because the sensor's life period is no longer valid. If the sensor has expired, replace it with a new one.
  • L
    Lisa Gomez MDAug 26, 2025
    What to do if no pump is running on Riken Keiki RX-415?
    • K
      Kristen PerryAug 26, 2025
      If the pump isn't running on your Riken Keiki Measuring Instruments and a comment is displayed, try re-operating the pump.

Summary

Precaution in Operation

Introduction to the RX-415

Safety Alert Symbols Definitions

Definitions of safety alert symbols used in the manual.

Safety Precautions for Operation

Instrument Components Overview

Instrument Operation Guide

Instrument Start-up Procedure

Step-by-step guide to power on and initialize the instrument.

Performing Gas Measurements

Performing gas concentration measurements with the instrument.

Alarm Functionality

Alarm Types and Functions

Details on different types of alarms and their behavior.

Instrument Maintenance Checks

Gas Calibration Process

Process for calibrating the instrument using calibration gases.

Troubleshooting Common Issues

Technical Specifications

Instrument Accessories

Riken Keiki RX-415 Specifications

General IconGeneral
Measuring Range0-100% LEL
Measurement Range0-100% LEL
Alarm Set Points20% LEL
Response TimeWithin 30 seconds
Power SourceRechargeable battery or AC adapter
Power SupplyAlkaline dry cell batteries (D size)
Output Signal4-20 mA analog output
DisplayDigital LCD
Sampling MethodPump suction
Operating Temperature Range-10°C to +40°C