Riken Keiki FI-8000 - User Manual

Riken Keiki FI-8000
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
PT3E-0525
Optical Interferometric Gas
Concentration Meter
FI-8000
Operating Manual
(PT3-052)

Questions and Answers

  • F
    feliciablakeSep 8, 2025
    What to do if key operations are disabled on Riken Keiki FI-8000 Measuring Instruments?
    • R
      Ryan Jacobs DDSSep 8, 2025
      If key operations are disabled on the Riken Keiki Measuring Instruments, it could be due to disturbances from static electricity. To resolve this, turn off the power, remove the battery unit, reinstall it in a non-hazardous area, and then turn the power back on.
  • B
    Brett JonesAug 23, 2025
    Why communication is not possible with Riken Keiki Measuring Instruments?
    • C
      chungmarcusAug 23, 2025
      If communication with your Riken Keiki Measuring Instruments fails, the main unit's position could be the problem. Try adjusting its position. Alternatively, another device using infrared (IR) might be interfering. Turn off other IR devices or prevent their interference to resolve this.
  • M
    matthew66Aug 25, 2025
    What to do if low battery voltage caution is displayed on Riken Keiki FI-8000?
    • E
      Eric LewisAug 25, 2025
      If the Riken Keiki Measuring Instruments displays '(CAUTION) LOW BATTEERY', it means the battery level is low. For an alkaline battery unit, turn off the power and replace the dry batteries with new ones in a non-hazardous area. If you're using a lithium-ion battery unit, turn off the power and charge the batteries in a non-hazardous area.
  • R
    rwhiteAug 27, 2025
    What to do if battery error is displayed on Riken Keiki FI-8000 Measuring Instruments?
    • T
      tericksonAug 27, 2025
      If the Riken Keiki Measuring Instruments displays '(ERROR) LOW BATTERY', it signifies that the battery level is too low to perform a measurement. If you have an alkaline battery unit, turn off the power and replace the dry batteries with new ones in a non-hazardous area. For lithium ion battery units, turn off the power and charge the batteries in a non-hazardous area.
  • C
    Carol BlakeAug 28, 2025
    Why my Riken Keiki Measuring Instruments showing incorrect reading?
    • M
      Mrs. Catherine PerkinsAug 28, 2025
      If your Riken Keiki Measuring Instruments shows incorrect readings, consider the following potential causes and solutions: * **Drifting sensor output:** Perform reference gas calibration. * **Incorrect gas selection:** Ensure you have selected the correct measuring and base gases. * **Interference gas:** Remove any interference gases present. * **Pressure issues:** Ensure GAS IN/GAS OUT is at atmospheric pressure. * **Tube issues:** Check for and fix any bended or clogged suction or exhaust tubes. * **Condensation:** Address any condensation inside the suction tube.
  • M
    Mario VasquezSep 22, 2025
    What does system error mean on Riken Keiki FI-8000?
    • A
      Adrienne DunnSep 22, 2025
      If the Riken Keiki Measuring Instruments displays '(ERROR) SYSTEM', it indicates an abnormality in the system. Request the dealer or Riken Keiki local representative for repair.
  • C
    Cody BishopSep 16, 2025
    What does UV intensity error mean on Riken Keiki FI-8000?
    • A
      Alan GardnerSep 16, 2025
      If the Riken Keiki Measuring Instruments displays '(ERROR) LOW BRIGHTNESS', it means that an abnormality occurred in the system. Request RIKEN KEIKI for the replacement or repair of interferometric sensor.
  • N
    Nathan MorrisonSep 18, 2025
    What does optical sensor unit error mean on Riken Keiki FI-8000?
    • C
      catherine47Sep 18, 2025
      If the Riken Keiki Measuring Instruments displays '(ERROR) OPTOCAL UNIT', it indicates an abnormality in communication signals from the optical sensor unit. Turn off and restart the concentration meter. If the symptom like this recurs frequently, the interferometric sensor needs to be repaired. Request RIKEN KEIKI for the repair.
  • G
    gholtSep 7, 2025
    What to do if Riken Keiki FI-8000 Measuring Instruments has abnormal operations?
    • H
      handersonSep 7, 2025
      If the Riken Keiki Measuring Instruments experiences abnormal operations, it might be due to disturbances by sudden static electricity noise. Try turning off and restarting the concentration meter.
  • N
    Nathan FriedmanSep 4, 2025
    Why the batteries cannot be charged in Riken Keiki FI-8000?
    • K
      Kerry WongSep 4, 2025
      If the Riken Keiki Measuring Instruments is not charging, first, ensure that the AC plug and DC plug of the AC powered charger are properly connected. If the issue persists, it may indicate a charging circuit abnormality, in which case you should request repair from the dealer or Riken Keiki local representative. Note that when fully charged batteries are charged again, the charging indicator lamp does not go on.

Summary

Important Safety Notices

2.1 Danger Cases

Details critical safety hazards and necessary countermeasures for hazardous areas.

2.2 Warning Cases

Outlines potential hazards and warnings for safe operation and handling.

2.3 Precautions

Provides essential precautions for correct and safe usage, including environmental limits.

2.4 Safety Information

Summarizes safety specifications, certifications, and critical warnings for use.

Product Components

3.1 Main Unit and Standard Accessories

Lists and describes the main unit and included standard accessories.

3.2 Names and Functions of Parts

Identifies and explains the function of each control and display element.

How to Use the FI-8000

4.2 Preparation for Start-up

Steps and precautions for preparing the meter, including battery and hose setup.

4.5 Measurement Method

Details on how to perform measurements using automatic or manual suction methods.

4.8 Reference Gas Calibration

Procedure for calibrating the meter using a specified reference gas.

Maintenance

6.2 Calibration

Recommended frequency and procedure for sensor span adjustment.

Troubleshooting Guide

Power On Issues

Addresses issues related to the unit failing to power on.

Reference Calibration Caution

Addresses caution messages related to reference gas calibration.

Battery Error

Actions to take when a battery error is indicated.

Incorrect Readings

Covers causes and actions for inaccurate measurement readings.

Product Specifications

9.1 Specification List

Comprehensive list of the FI-8000's technical specifications.

Riken Keiki FI-8000 Specifications

General IconGeneral
BrandRiken Keiki
ModelFI-8000
CategoryMeasuring Instruments
LanguageEnglish