PT2E-1601
Smart transmitter/
Gas Detector Head
SD-1
Operating Manual
(PT2-160)
Request for the Customers
Read and understand this operating manual before using the detector.
Use the detector in accordance with the operating manual.
Regardless of warranty period, we shall not make any compensation for accidents and
damage caused by using this product.
The compensation shall be made only under the warranty policy of products or parts
replacement.
Because this is a safety unit, a regular maintenance for every six months and daily
maintenance must be performed.
If any abnormality is found in the gas monitor, notify it to RIKEN KEIKI immediately.
(Visit our Web site to find your nearest RIKEN KEIKI office.)

Questions and Answers

  • B
    Benjamin GordonSep 23, 2025
    Why is my Riken Keiki Other responding slowly?
    • R
      Rebecca ShermanSep 23, 2025
      A slow response on your Riken Keiki Other device indicates that the sensor sensitivity may have deteriorated. Replace the sensor with a new one.
  • R
    Richard MartinAug 23, 2025
    What to do if Riken Keiki SD-1 alarm with no gas leak?
    • P
      patricia57Aug 23, 2025
      If your Riken Keiki Gas Detectors trigger a gas alarm when there's no leak, here's what might be happening: * **Interference Gas:** The detector may be reacting to other gases, including solvents, so an interference gas test should be conducted. * **Noise Disturbance:** External noise can cause false alarms. Try turning the detector off and on again. If this happens frequently, take steps to reduce noise. * **Sudden Environmental Changes:** Rapid changes in temperature can affect the detector's readings. Environmental changes should be eliminated.
  • J
    Jessica WarrenAug 25, 2025
    Why does the reading on Riken Keiki SD-1 keep rising or dropping?
    • J
      Julie SpencerAug 26, 2025
      If the reading on your Riken Keiki Gas Detectors rises or drops and stays that way, it could be due to several reasons. One possibility is drifting of the sensor output, in which case you should perform a zero adjustment. Another reason could be the presence of interference gas, which requires an interference gas test. A slow leak of the gas to be detected may also be the reason. Environmental changes can also cause this issue, so performing a zero adjustment (AIR adjustment) may help.
  • S
    sullivanstephenAug 27, 2025
    How to fix sensor abnormalities E-1 on Riken Keiki Gas Detectors?
    • D
      dpenaAug 27, 2025
      If your Riken Keiki Gas Detectors display a 'Sensor abnormalities E-1' error, it could be due to a few reasons. First, check that the sensor cable is securely fastened to the terminal plate. It could also be due to zero drift caused by environmental changes or aging. In this case, perform a zero adjustment. If the problem continues after the zero adjustment, the sensor may be faulty and needs replacement.
  • D
    David JacksonAug 27, 2025
    Why is sensitivity calibration impossible on Riken Keiki Gas Detectors?
    • M
      margaretchaseAug 27, 2025
      If you're unable to calibrate the sensitivity of your Riken Keiki Gas Detectors, it could be due to an improper calibration gas concentration. Make sure you are using the correct calibration gas. Another potential cause is deteriorated sensor sensitivity, which would require replacing the sensor.
  • D
    Daniel RomanSep 16, 2025
    What to do if Riken Keiki Other is showing abnormal operations?
    • B
      brittany86Sep 16, 2025
      If your Riken Keiki Other device is exhibiting abnormal operations, it might be due to disturbances like sudden surge noise. Try turning off and restarting the detector. If the issue persists, take appropriate measures to eliminate the noise.
  • N
    Nathan SandovalSep 13, 2025
    How to fix 'Sensor abnormalities E-1' on Riken Keiki Other?
    • M
      Michael PopeSep 13, 2025
      If your Riken Keiki Other device displays a 'Sensor abnormalities E-1' error, check if the sensor cable is securely fastened to the terminal plate. The error may also appear because zero drift caused by environmental changes or aging deterioration is out of the range of zero follower; in this case, perform a zero adjustment. If the symptom persists after the zero adjustment, replace the sensor with a new one. The sensor itself may be faulty, requiring replacement.
  • W
    william13Sep 14, 2025
    Why does the reading on my Riken Keiki Other keep rising or dropping?
    • C
      Christina GreenSep 15, 2025
      If the reading on your Riken Keiki Other device rises or drops and remains that way, first perform a zero adjustment, including an AIR adjustment if environmental changes are suspected. This issue might be caused by drifting of sensor output, environmental changes, slow leak. In the case of a slow leak, take a remedial measure, i.e., taking actions the same as those for the gas alarm.
  • H
    hdickersonSep 5, 2025
    What to do if Riken Keiki Gas Detectors trigger a gas alarm despite of no gas leak?
    • L
      Luke WilliamsSep 6, 2025
      If your Riken Keiki Gas Detectors trigger a gas alarm without a gas leak or other abnormalities, try turning the detector off and restarting it. If the issue persists, take measures to eliminate any noise disturbances. If a sudden change in the environment is the cause, take preventive actions to eliminate them.
  • C
    Christopher TaylorSep 3, 2025
    What to do if Riken Keiki SD-1 display sensor abnormalities E-1?
    • K
      Kyle VazquezSep 3, 2025
      If your Riken Keiki Gas Detectors are displaying sensor abnormalities (E-1), verify the sensor cable is securely fastened to the terminal plate. Perform a zero adjustment; if the problem continues, replace the sensor. If zero drift caused by environmental changes or aging deterioration is out of the range of zero follower, replace the sensor with a new one.

Summary

Operating Precautions

Outline of the Product

Definition of DANGER, WARNING, CAUTION, and NOTE

Defines hazard symbols and their meanings.

Important Notices on Safety

Danger cases

Lists critical safety warnings related to explosion-proof construction.

Warning cases

Details critical warnings for power, grounding, and protective functions.

Precautions

Lists important precautions for radio interference, restart, and maintenance.

Safety information

Provides technical data and installation diagrams for explosion-proof construction.

How to Use

Precautions for installation points

Lists environmental and physical precautions for installing the detector.

Precautions for system designing

Covers power supply stability, noise, and lightning protection for system design.

How to Operate

Operations and Functions

Maintenance

Maintenance intervals and items

Outlines daily, monthly, and regular maintenance tasks and frequencies.

Regular maintenance mode

Details the different modes for regular maintenance procedures.

Gas calibration method

Explains the process of zero and span calibration.

Troubleshooting

Abnormalities on Unit

Lists unit-level symptoms, causes, and corrective actions.

Abnormalities of Readings

Lists reading-related symptoms, causes, and corrective actions.

Riken Keiki SD-1 Specifications