M2A Transmitter Operator’s Manual 21
OLED Display
The OLED display is located at the top of the control PCB. It indicates the current gas reading and displays
messages and parameters in the M2A’s programs.
Control Buttons
The M2A includes three push button switches that allow you to enter the M2A’s operating modes, navigate
through the modes, update settings, and save changes to the settings. The push button switches are located along
the bottom edge of the control PCB (see Figure 12). The UP/YES button is on the left, the DOWN/NO button is
in the middle, and the ENTER button is on the right.
Just above each push button switch is a magnetic switch with the same function as the push button switch below
it. The magnetic switches are for use in non-intrusive calibration. They are actuated by bringing the magnetic
wand close enough to them to actuate them. Although the magnetic switches have the same functions as the push
button switches, it is not practical to use them for operations other than calibration because it is not possible to
actuate two magnetic switches at once with only one magnetic wand. Since displaying the Information Screen
only requires the use of one switch, the wand may be used to show the Information Screen (see “Information
Screen” on page 37).
Status LEDs
The M2A includes five status LEDs that are located above the display (see Figure 12).
Fail LED
The fail LED turns on when the M2A is experiencing a fail condition. A fail condition can be caused by a
detector failure or low detector signal.
Alarm 1 LED
The alarm 1 LED is on when the M2A is experiencing an alarm 1 condition.
Alarm 2 LED
The alarm 2 LED is on when the M2A is experiencing an alarm 2 condition.
RX & TX LED’s
These LED’s indicate data being received (RX) and transmitted (TX) when the M2A’s Modbus output is
operating.
Table 8: M2A Control Button Functions
Switch Function
UP (YES) Saves settings
Changes the displayed setting
Enters the Calibration Program
Enters Gas Type Mode (press with DOWN/NO button)
Enters Configuration Mode (press with ENTER button)
DOWN (NO) Cancels setting changes
Changes the displayed setting
Enters Gas Type Mode (press with UP/YES button)
Enters Modbus Mode (press with ENTER button)
Displays the Information Screen
ENTER Initiates operations
Enters Configuration Mode (press with UP/YES button)
Enters Modbus Mode (press with DOWN/NO button)
Functions as an alarm reset switch