8 M2A Transmitter Operator’s Manual
Alarm Settings
(Alarm 1/Alarm 2)*,
cont’d
Carbon Monoxide (all ranges): 25 ppm / 50 ppm
Chlorine (0 - 3.00 ppm): 0.50 ppm / 1.00 ppm
Chlorine (0 - 5.00 ppm): 0.50 ppm / 1.50 ppm
Chlorine (0 - 10.0 ppm): 1.0 ppm / 3.0 ppm
Chlorine Dioxide (0 - 1.00 ppm): 0.10 ppm / 0.30 ppm
Chlorine Dioxide (0 - 3.00 ppm): 0.30 ppm / 1.00 ppm
Chlorine Dioxide (0 - 5.00 ppm): 0.50 ppm / 1.00 ppm
Combustible Gas (%LEL): 10 %LEL / 50 %LEL
Combustible Gas (0-100%Volume): 100 % / 100 %
Hydrogen Cyanide: 5.0 ppm / 10.0 ppm
Hydrogen Sulfide (0 - 50 ppm): 5.0 ppm / 10.0 ppm
Hydrogen Sulfide (0 - 100 ppm): 10 ppm / 50 ppm
Methane (ppm): 5000 ppm / 8000 ppm
Methane (0-5.00%Volume): 0.5% / 2.5 %
Oxygen: 19.5% (decreasing) / 23.5% (increasing)
Phosphine: 0.10 ppm / 0.30 ppm
Sulphur Dioxide: 1.00 ppm / 2.00 ppm
Construction (housing) Explosion-proof Junction Box, NEMA 4X
Area Classification Explosion-proof for Class I, Groups B, C, and D (Combustible, CO
2
, CSA Type CO and H
2
S,
and CSA Type oxygen)
Sampling Method Diffusion
Input Power 10 - 30 VDC
Controls Three push button switches
Three magnetic switches for non-intrusive calibration
Weight 4.5 lbs.
Signal Output 4 to 20 mA with impedance maximums listed below
-12 VDC input: 500 ohms impedance max
-24 VDC input: 1000 ohms impedance max
RS-485 Modbus
Explosion Proof Non-Explosion Proof
Operating Temperature
Catalytic LEL/ppm -40°C to 75°C N/A
Oxygen -20°C to 45°C -20°C to 50°C
H
2
S -40°C to 40°C -40°C to 50°C
CO -5°C to 40°C -20°C to 50°C
Toxic (ESM-01) N/A -10°C to 40°C
Tox ic (CT-7 Series) N/A -20°C to 40°C
Infrared Combustible -20°C to 50°C -40°C to 50°C
Infrared CO
2
-20°C to 50°C -40°C to 50°C
Table 1: M2A Specifications