Polilight Flare Plus 2 manual –
17
Table 1 shows the settings needed for certain crime scene examinations.
Table 1 – Examination and recording conditions
Technique Mode Polilight Band Barrier Filter
Ninhydrin developed
fingerprints
Absorption White light
BP565
Ninhydrin developed
and cadmium post-
treated fingerprints
Absorption 505 nm
Darkened room, no
barrier filter
Photoluminescence
(77K)
505 nm OG550 or BP590
Ninhydrin developed
and zinc post-treated
fingerprints
Absorption
490 nm or 505 nm
Darkened room no
barrier filter
Photoluminescence
(77K)
OG550 or BP565
DFO developed
fingerprints
Photoluminescence (RT)
545 nm BP600 or BP610
530 nm OG590 or BP590
505 nm OG550 or BP565
IND-Zn developed
fingerprints
Absorption White light BP530
Photoluminescence (RT) 505 nm OG550 or BP565
Cyanoacrylate
developed and
rhodamine 6G stained
fingermarks
Photoluminescence
530 nm OG590 or BP590
505 nm OG550 or BP565
Cyanoacrylate
developed and basic
yellow 40 stained
fingermarks
Photoluminescence 450 nm
OG515, OG550 or
BP565
Blood stains (or
fingerprints in blood)
Absorption
415 nm
Dark conditions, no
barrier filter
415 nm or White Light BP415
850 nm Infra Red IR senstive Camera
Semen stains Photoluminescence
UV or 415 nm Yellow Goggles
440 or 450 or 505 nm Orange Goggles
530 nm Red Goggles
LIGHT APPLICATIONS