EasyManua.ls Logo

Rohde & Schwarz CMW500 Series - Page 104

Rohde & Schwarz CMW500 Series
291 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Basic Instrument Functions
R&S
®
CMW500
104User Manual 1173.9463.02 ─ 02
RF Loop Test
A signal which is generated on one of the installed TRX modules is routed to a frontend
connector RF <n> COM and back to the TRX module. The main purpose of the test is to
verify the connections between the TRX module and the frontend. The RF loop is mea-
sured at different frequencies and TX levels using an additional BB measurement module.
The header row of the overall loop test shows the output levels in dBm at the RF <n>
COM connectors. The actual levels measured are approx. 6 dB above the equivalent RF
<n> COM levels.
Sample Bus Test
A pseudo-random bit sequence (PRBS) is transferred over the sample bus between two
hardware modules. The R&S CMW500 compares the transmitted PRBS with the received
PRBS and verifies that no bit errors occur in the transmission path.
External Tests
One or more R&S CMW500 connectors are tested using external cable connections.
External tests require the "User (Extended)" mode; see "User Mode" on page 95.
Note: The "IP Access Test External" shown above requires a fixed IP address 172.22.1.3
for the LAN REMOTE network adapter; see chapter 2.7.1.2, "Applications with Fixed IP
Addresses", on page 33.
When the test is started, the R&S CMW500 prompts for the appropriate cable connection,
e.g.:
A similar message ("Please disconnect...") after the test is finished.
Setup Dialog

Table of Contents

Related product manuals