Remote Control
R&S
®
CMW500
151User Manual 1173.9463.02 ─ 02
Signaling generator control commands are of the following type (see also Firmware
Applications):
SCPI subsystem <Application>, e.g. Generator instance State
SOURce :GSM :SIGNaling<i> :CELL:STATe ON | OFF
Example: SOURCe:GSM:SIGNaling:CELL:STATe ON | OFF
The :CELL mnemonic is omitted in some signaling applications.
SOURce:<Application>:SIGNaling<i>:CELL:STATe ON
Starts the signaling generator, reserves all necessary hardware and system resources
and changes to the generator state "PENDing", then "ON". If the cell signal is already
turned on the command has no effect.
If the hardware and system resources are already assigned to another firmware appli-
cation, this firmware application is released in order to start the signaling generator; see
Resource and Path Management.
If the signaling generator cannot be started due to an unrecoverable resource conflict
(e.g. a missing software option) it remains in the OFF state. The SCPI error "-213, Init
ignored, ...", is generated. See also Causes for Task Conflicts.