Command Reference
R&S
®
CMW500
227User Manual 1173.9463.02 ─ 02
Example:
STAT:OPER:TASK:A:GPRF:MEAS:POW:PTR 0
STAT:OPER:TASK:A:GPRF:MEAS:POW:NTR 65535
Configure the transition registers so that the EVENt part of the
status register reports only transitions from 1 to 0.
INIT:GPRF:MEAS:POW
STAT:OPER:TASK:A:GPRF:MEAS:POW:WEV? 4,1000
Initiate a GPRF power measurement and wait until it has left the
state RUN (bit number 2, decimal 4). The timeout is set to 1000
ms.
The returned value is 0 or 4 (timeout occurred or state RUN left).
Usage: Query only
Firmware/Software: V2.0.10
STATus:
OPERation:BIT<n>:CONDition?
STATus:OPERation:BIT<n>[:EVENt]?
Returns bit no. <n> of the CONDition or EVENt part of the STATus:OPERation register,
see also Structure of an SCPI Status Register. To return the entire parts see STATus:
OPERation:CONDition and STATus:OPERation[:EVENt].
Suffix:
<n>
.
8 to 12
Number of the bit
Return values:
<BitValue> 0 | 1
Example:
STAT:OPER:BIT9:COND?
Query bit no. 9 of the CONDition part of the STATus:OPERation
register.
Usage: Query only
Firmware/Software: V1.0.4.11
STATus:OPERation:BIT<n>:ENABle <BitValue>
STATus:OPERation:BIT<n>:NTRansition <BitValue>
STATus:OPERation:BIT<n>:PTRansition <BitValue>
Sets bit no. <n> of the ENABle, NTRansition or PTRansition part of the STATus:OPER-
ation register, see also Structure of an SCPI Status Register. To set the entire parts see
STATus:OPERation:ENABle, STATus:OPERation:NTRansition and STATus:
OPERation:PTRansition.
Suffix:
<n>
.
8 to 12
Number of the bit
Parameters:
<BitValue> 0 | 1
Instrument-Control Commands