EasyManua.ls Logo

Rohde & Schwarz CMW500 Series - Page 248

Rohde & Schwarz CMW500 Series
291 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
CMW500
248User Manual 1173.9463.02 ─ 02
SYSTem:COMMunicate:SOCKet<i>:VRESource?
Queries the VISA resource string of the socket resource (direct socket communication);
see Remote Control Operation.
Suffix:
<i>
.
1 to 2
Selects the remote channel.
Return values:
'<String>' VISA address string; see example below.
*RST: *RST has no effect on the value. The address string
depends on the host name and on the port number;
see example below.
Example:
SYSTem:COMMunicate:NET:HOSTname
'hh346999.domain.net'
SYSTem:COMMunicate:SOCKet2:PORT 1500
SYSTem:COMMunicate:SOCKet2:VRESource?
Define a host name, set a port and query the VISA resource string
of the socket resource (channel 2). The response is
'TCPIP::hh346999.domain.net::1500::SOCKET'.
Usage: Query only
Firmware/Software: V1.0.5.3
SYSTem:COMMunicate:GPIB<i>[:SELF]:ADDR <AddressNo>
Sets the primary GPIB address of the analyzer.
Suffix:
<i>
.
1 to 4, depending on installed options
Selects the remote channel.
Parameters:
<AddressNo> GPIB address; integer number
Range: 0 to 30
*RST: a *RST has no effect on the value
Example:
SYSTem:COMMunicate:GPIB:ADDRess 10
Set the GPIB address to 10 (channel 1).
Firmware/Software: V1.0.0.4
SYSTem:COMMunicate:GPIB<i>[:SELF]:ENABle <State>
Enables or disables the GPIB interface.
Suffix:
<i>
.
1 to 4, depending on installed options
Selects the remote channel.
Instrument-Control Commands

Table of Contents