Command Reference
R&S
®
CMW500
248User Manual 1173.9463.02 ─ 02
SYSTem:COMMunicate:SOCKet<i>:VRESource?
Queries the VISA resource string of the socket resource (direct socket communication);
see Remote Control Operation.
Suffix:
<i>
.
1 to 2
Selects the remote channel.
Return values:
'<String>' VISA address string; see example below.
*RST: *RST has no effect on the value. The address string
depends on the host name and on the port number;
see example below.
Example:
SYSTem:COMMunicate:NET:HOSTname
'hh346999.domain.net'
SYSTem:COMMunicate:SOCKet2:PORT 1500
SYSTem:COMMunicate:SOCKet2:VRESource?
Define a host name, set a port and query the VISA resource string
of the socket resource (channel 2). The response is
'TCPIP::hh346999.domain.net::1500::SOCKET'.
Usage: Query only
Firmware/Software: V1.0.5.3
SYSTem:COMMunicate:GPIB<i>[:SELF]:ADDR <AddressNo>
Sets the primary GPIB address of the analyzer.
Suffix:
<i>
.
1 to 4, depending on installed options
Selects the remote channel.
Parameters:
<AddressNo> GPIB address; integer number
Range: 0 to 30
*RST: a *RST has no effect on the value
Example:
SYSTem:COMMunicate:GPIB:ADDRess 10
Set the GPIB address to 10 (channel 1).
Firmware/Software: V1.0.0.4
SYSTem:COMMunicate:GPIB<i>[:SELF]:ENABle <State>
Enables or disables the GPIB interface.
Suffix:
<i>
.
1 to 4, depending on installed options
Selects the remote channel.
Instrument-Control Commands