Basic Instrument Functions
R&S
®
CMW500
85User Manual 1173.9463.02 ─ 02
Fig. 5-2: Dialog for 1 sub-instrument
Fig. 5-3: Dialog for several sub-instruments
The reset/preset can be performed for the entire R&S CMW500 or it can be limited to the
current firmware application.
If the R&S CMW500 has been split into sub-instruments (see Instrument Setup Dialog),
the reset/preset can also be limited to the current sub-instrument.
In addition to reset state and preset state the instrument supports also a factory default
state. All three states are described below.
Preset
Sets the instrument parameters to values suitable for local/manual interaction. In partic-
ular, the Preset state comprises the following settings:
All measurements are repeated continuously ("Repetition: Continuous").
The R&S CMW500 uses long statistics cycles (for reliable statistical evaluations).
The preferred modes and features of the current software version are active. Con-
sequently, the "Preset" state may change after a software upgrade.
Exceptions:
The following R&S CMW500 settings are not affected by Preset:
Address information assigned to the instrument (e.g. the IP address)
Utility Dialogs