V L F > H F R E C E I V E R S
f
E K 8 9 5 / E K 8 9 6
Operator Manual
f
List of Commands
6075.6991.82.01
> 2.8 >
2.6 List of Commands
Code Function (entry in) Entry
Output
(requested with ?)
Remark
A Address 00 to 99 01 to 99 see A3.8.1
1)
BI Initiate BIT
Inquire BIT status
s
0= no fault
10= synthesizer
missing
1= synthesizer
defective
20= HF unit
missing
2= HF unit
defective
5= module bus
defective
60= IF / AF proces>
sor missing
6= IF / AF proces>
sor defective
7= IF Converter
UX 895
defective
see A3.7.4
1)
C Inquire CM status 0= no fault
1= synthesizer
defective
4= IF / AF proces>
sor defective
5= synthesizer
and IF / AF
processor
defective
see A3.7.5
1)
CL Inhibit channel 0 to 999 see A3.7.11
1)
CLa Inhibit all channels see A3.7.11
1)
ERR Error status 0,0= no fault see A3.8.8
1)
F Frequency (Hz) 0 to 30000000 0 to 30000000 see 2.3.1
FIB Inquire IF filter bandwidths 1= 150 Hz
3= 300 Hz
6= 600 Hz
10= 1000 Hz
15= 1500 Hz
21= 2100 Hz
see A3.8.3
1)
1)
These commands are described in the User Manual 6045.6712.12.xx.