Remote Commands in the Receiver Application
R&S
®
ESW
462User Manual 1177.6298.02 ─ 08
Parameters:
<Measurements> <numeric value> (integer only)
Number of measurements considered in calculating an average
trace.
When you set a count of 1, the R&S ESW performs a single
measurement over the selected scan range.
Range: 0 to 32767
*RST: 0
Example: //Define 10 measurements in single measurement mode
SWE:COUN 10
INIT:CONT OFF
//Start sweep and wait for its end
INIT;*WAI
Manual operation: See "Scan Count" on page 113
[SENSe:]SWEep:SPACing <StepMode>
This command selects the frequency step mode.
Note that the command has no effect on the scale and display of the frequency axis.
Parameters:
<StepMode> LINear
Linear frequency steps with a fix stepsize.
LOGarithmic
Logarithmic frequency steps with the stepsize being a percent-
age of the current frequency.
AUTO
The stepsize is coupled to the resolution bandwidth to get the
best measurement results.
*RST: AUTO
Example: //Select logarithmic frequency steps
SWE:SPAC LOG
Manual operation: See "Selecting the frequency step mode" on page 118
15.5.7 Peak Search
Commands to configure the peak search described elsewhere.
●
FORMat:DEXPort:DSEParator on page 581
CALCulate<n>:MARKer<m>:PEXCursion.........................................................................463
CALCulate<n>:PSEarch:ADD..........................................................................................463
CALCulate<n>:PEAKsearch:ADD....................................................................................463
CALCulate<n>:PSEarch:AUTO....................................................................................... 464
CALCulate<n>:PEAKsearch:AUTO..................................................................................464
CALCulate<n>:PSEarch:CLEar[:IMMediate]..................................................................... 464