Remote Commands in the Receiver Application
R&S
®
ESW
564User Manual 1177.6298.02 ─ 08
Manual operation: See "Visibility" on page 247
CALCulate<n>:LIMit<li>:CONTrol:OFFSet <Offset>
This command defines an offset for a complete limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<Offset> Numeric value.
The unit depends on the scale of the x-axis.
*RST: 0
Default unit: HZ
Manual operation: See "X-Offset" on page 247
CALCulate<n>:LIMit<li>:COPY <Line>
This command copies a limit line.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<Line> 1 to 8
number of the new limit line
<name>
String containing the name of the limit line.
Example:
CALC:LIM1:COPY 2
Copies limit line 1 to line 2.
CALC:LIM1:COPY 'FM2'
Copies limit line 1 to a new line named FM2.
Manual operation: See "Copy Line" on page 248
CALCulate<n>:LIMit<li>:DELete
This command deletes a limit line.
Suffix:
<n>
.
Window
<li> Limit line
Manual operation: See "Delete Line" on page 248