Remote Commands in the Receiver Application
R&S
®
ESW
567User Manual 1177.6298.02 ─ 08
ON | 1
Switches the function on
Example:
CALC:LIM3:TRAC2:CHEC ON
Switches on the limit check for limit line 3 on trace 2.
Manual operation: See "Traces to be Checked" on page 247
CALCulate<n>:LIMit<li>:UPPer:OFFSet <Offset>
This command defines an offset for a complete upper limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<Offset> Numeric value.
*RST: 0
Default unit: dB
Manual operation: See "Y-Offset" on page 248
CALCulate<n>:LIMit<li>:UPPer:STATe <State>
This command turns an upper limit line on and off.
Before you can use the command, you have to select a limit line with CALCulate<n>:
LIMit<li>:NAME on page 573.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation: See "Visibility" on page 247
MMEMory:LOAD<n>:LIMit <FileName>
Loads the limit line from the selected file in .CSV format.
Suffix:
<n>
.
irrelevant