Remote Commands in the Receiver Application
R&S
®
ESW
652User Manual 1177.6298.02 ─ 08
(➙ 10 dB Minimum Attenuation)
Parameters:
<State> ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: NONE
Example:
INST:COUP:PROT ALL
Synchronizes the "10 dB Min" parameter.
Manual operation: See "Synchronizing parameters across all measurement chan-
nels" on page 348
INSTrument:COUPle:SPAN
<State>
This command turns synchronization of the start and stop frequency between mea-
surement channels on and off.
Parameters:
<State> ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: NONE
Example:
INST:COUP:SPAN ALL
Synchronizes the start and stop frequency.
Manual operation: See "Synchronizing parameters across all measurement chan-
nels" on page 348
INSTrument:COUPle:VBW <State>
This command turns synchronization of the video bandwidth between measurement
channels on and off.
Parameters:
<State> ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: NONE
Example:
INST:COUP:VBW ALL
Synchronizes the video bandwidth.
Manual operation: See "Synchronizing parameters across all measurement chan-
nels" on page 348
General Instrument Setup