Measurement Equipment and Accessories R&S FSH
1145.5973.82 2.6 E-11
Measurement Equipment and Accessories
Item Type of
equipment
Specifications recommended Equipment
recommended
R&S order
no.
Use
1 Signal generator Frequency: 10 MHz to 3 GHz
Uncertainty of frequency: 0.1 ppm
Phase noise at 500 MHz:
< -100 dBc (1Hz) @ 10 kHz
< -110 dBc (1Hz) @ 100 kHz
< -130 dBc (1Hz) @ 1 MHz
R&S SML Frequency accuracy of
reference oscillator
2 6-dB divider
(power splitter)
Frequency response
3 Power meter R&S NRVD 0857.8008.02 Frequency response
4 Power sensor 1 MHz to 3 GHz
RSS 0.8%
Meter noise 20 pW
R&S NRV-Z4 0828.3618.02 Frequency response
Adjusting the reference frequency accuracy
Test equipment Signal generator (refer to section "Measurement Equipment and
Accessories", item 2)
Test setup Connect the generator to the RF input.
Generator settings
- Frequency 1 GHz
- Level -30 dBm
R&S FSH settings - [ PRESET ]
- [ FREQ : 1 GHz ]
- [ SPAN : 100 kHz ]
- [ BW : MANUAL RES BW : 10 kHz ]
- [ MARKER : MARKER MODE : FREQ COUNT ]
Adjustment The program will adjust the reference oscillator to 10 MHz ± 1 ppm.