Remote control commands
R&S
®
RTA4000
436User Manual 1335.7898.02 ─ 08
CHANnel<m>:STATe <State>
Switches the channel signal on or off.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<State> ON | OFF
CHANnel<m>:AON
Switches all analog channels on.
Suffix:
<m>
.
The suffix is irrelevant.
Usage: Event
CHANnel<m>:AOFF
Switches all analog channels off.
Suffix:
<m>
.
The suffix is irrelevant.
Usage: Event
CHANnel<m>:SCALe <Scale>
Sets the vertical scale for the indicated channel.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Scale> Scale value, given in Volts per division.
Range: 1e-3 to 10 (without probe attenuation)
*RST: 5e-3
Default unit: V/div
CHANnel<m>:RANGe <Range>
Sets the voltage range across the all vertical divisions of the diagram. Use the com-
mand alternatively instead of CHANnel<m>:SCALe.