Remote control commands
R&S
®
RTA4000
527User Manual 1335.7898.02 ─ 08
MEASurement<m>:GATE <State>
Activates or deactivates the measurement gate.
To set the gate, use:
MEASurement<m>:GATE:MODE and
MEASurement<m>:GATE:ABSolute:STARt and MEASurement<m>:GATE:
ABSolute:STOP
or MEASurement<m>:GATE:RELative:STARt and MEASurement<m>:GATE:
RELative:STOP
Suffix:
<m>
.
1..8
The suffix is irrelevant. One gate is used for all measurement
places.
Parameters:
<State> ON | OFF
*RST: OFF
MEASurement<m>:GATE:MODE <GateMode>
Defines whether the gate is defined in absolute or relative values.
Suffix:
<m>
.
1..4
The suffix is irrelevant. One gate is used for all measurement
places.
Parameters:
<GateMode> RELative | ABSolute
*RST: REL
MEASurement<m>:GATE:ABSolute:STARt <StartTime>
Defines the absolute start time for the measurement gate.
Suffix:
<m>
.
1..4
The suffix is irrelevant. One gate is used for all measurements.
Parameters:
<StartTime> *RST: -300e-6
Default unit: s
MEASurement<m>:GATE:ABSolute:STOP <StopTime>
Defines the absolute end time for the measurement gate.
Measurements