Remote control commands
R&S
®
RTA4000
790User Manual 1335.7898.02 ─ 08
Parameters:
<PatternPeriod> Numeric value (Period = Pattern length * Bit time)
Range: MIN 1 Sample * 20ns = 20ns to MAX 2048 Sam-
ples * 42s = 10416s (approx. 2.89h)
*RST: 2.000E-06
Default unit: s
PGENerator:PATTern:FREQuency <PatternFrequency>
Defines the frequency (period) value of the pattern generator function.
Parameters:
<PatternFrequency> Numeric value
Default unit: Hz
17.14.2.3 Counter pattern
PGENerator:PATTern:COUNter:FREQuency.....................................................................790
PGENerator:PATTern:COUNter:DIRection........................................................................ 790
PGENerator:PATTern:COUNter:FREQuency <Period>
Defines the frequency value of the pattern generator counter function. The user fre-
quency always refers to the switching of the pattern condition. This results in square
waveforms for individual pins.
Pin Frequency
S0 f/2
S1 f/4
S2 f/8
S3 f/16
Parameters:
<Period> Numeric value
Range: 2.380952425301E-02 to 2.500000000000E+07
*RST: 1.000000000000E+05
Default unit: Hz
PGENerator:PATTern:COUNter:DIRection <CountDirection>
Sets the pattern generator counter direction.
Parameters:
<CountDirection> UPWard | DOWNward
*RST: UPW